- at distances between 0 and 1 cm, the settings had changed in 58 percent of the valves.
- after exposure at distances between 1 and 2.5 cm the settings had changed in 5 percent of valves.
- after exposure at distances between 2.5 and 5 cm the settings had changed in only 1 percent of valves.
- no changes in valve settings were identified after exposures at higher distances.
- after exposure of programmable valves to the iPad 2 without a cover, which was only tested at distances between 0 and 1 cm, the researchers found that the settings had changed in 67 percent of the valves.
Reference: Medical Express. Tablet computers may interfere with settings on magnetically programmable shunt valves. Retrieved from the web on 6/27/2012 at http://medicalxpress.com/news/2012-06-tablet-magnetically-programmable-shunt-valves.html